Desarrollo de un dispositivo de medición usando un Geiger Müller para la determinación de actividad en un prototipo de generador de Tc-99m
dc.contributor.author | Urquizo, Rafael | |
dc.contributor.author | Gago, Javier | |
dc.contributor.author | Cruz-Saco, César | |
dc.contributor.author | Rojas, Jorge | |
dc.contributor.author | Mendoza, Pablo | |
dc.creator | Rojas, Jorge | |
dc.creator | Urquizo, Rafael | |
dc.creator | Mendoza, Pablo | |
dc.creator | Cruz-Saco, César | |
dc.creator | Gago, Javier | |
dc.date.accessioned | 2016-11-04T14:54:26Z | |
dc.date.available | 2016-11-04T14:54:26Z | |
dc.date.issued | 2016-05 | |
dc.identifier.citation | Urquizo R, Gago J, Cruz-Saco C, Rojas J, Mendoza P. Desarrollo de un dispositivo de medición usando un Geiger Müller para la determinación de actividad en un prototipo de generador de Tc-99m. Informe Científico Tecnológico. 2014; 14:133-136. | es_PE |
dc.identifier.issn | 1684-1662 | |
dc.identifier.uri | https://hdl.handle.net/20.500.13054/609 | |
dc.description.abstract | En el presente artículo se muestra la implementación de un sistema de medición empleando un detector Geiger Müller (GM) con la finalidad de adaptarlo a un prototipo de generador de Tc-99m. La señal de respuesta del sistema de medición diseñado en términos de tasa de conteo es linealmente proporcional a la variación de la actividad permitiendo medir entre 280 mCi y 170 mCi de Tc-99m con un error relativo del ± 2,8 %. No obstante, queda pendiente realizar mayores pruebas a fin de evaluar su correlación a un nivel de actividad menor a 20 mCi con el fin de obtener un rango adecuado de uso. | es_PE |
dc.description.abstract | This article presents the implementation of a measurement system using a Geiger Müller detector (GM) in order to adapt it into a Tc-99m generator prototype. The response signal of the measurement system designed in terms of count rate is linearly proportional to the variation of the activity measured between 280 and 170 mCi of Tc-99m with a relative error of ± 2,8 %. However, further tests are needed to evaluate the correlation for an activity level lower than 20 mCi in order to obtain an adequate range of use. | |
dc.description.sponsorship | INNOVATE-PERU contrato FINCyT N° 162-2013. | |
dc.format | application/pdf | es_PE |
dc.language.iso | spa | es_PE |
dc.publisher | Instituto Peruano de Energía Nuclear | es_PE |
dc.rights | info:eu-repo/semantics/openAccess | es_PE |
dc.source | Repositorio Institucional del Instituto Peruano de Energía Nuclear | es_PE |
dc.source | Instituto Peruano de Energía Nuclear | es_PE |
dc.subject | Generadores de radioisótopos | es_PE |
dc.subject | Tecnecio 99 | es_PE |
dc.subject | Prototipos | es_PE |
dc.subject | Técnicas de conteo | es_PE |
dc.title | Desarrollo de un dispositivo de medición usando un Geiger Müller para la determinación de actividad en un prototipo de generador de Tc-99m | es_PE |
dc.title.alternative | Development of a measurement device, using a Geiger Müller type detector, for the determination of the activity in a Tc-99m generator prototype | es_PE |
dc.type | info:eu-repo/semantics/article | es_PE |
dc.publisher.country | PE | es_PE |